What should I do if the electromagnetic vibration table fails? Qinsun

2022-08-16

The electromagnetic vibration platform simulates the structural strength, material wear, the standard value of the components, the poor contact of the component, the short circuit of the circuit, and the disconnected instability of the component, and the adverse parts will be screened early.

The electromagnetic vibration platform cannot connect to the computer control (with microcomputer model)

1. Check whether the line around the computer is correctly connected, please connect and confirm it again
2. Check whether the RS232 terminal is connected and contact well, and it needs to be re -connected
3. Check whether the power switch of the vibration platform control box is turned on. ON (that is, turn on the vibration platform power supply)
4. Then check the RS232 serial computer setting is COM2 or COM1, just replace the settings
5. Whether to open the program screen correctly, whether Correct settings, refer to the setting of the computer operation manual


The electromagnetic vibration table cannot be stopped after the setting time

1. Use the meat to see the abnormal noise of the test object tied the contact surface. Whether the spring steel sheet inside the table is broken. If the size (long and wide) is measured (long and wide), the manufacturer can be sent to your company
2. Some numbers appear in the computer program and prompt the contact manufacturer to provide the authorization code:
3. Check whether the computer is working normally (not working on the vibration platform)
4. Confirm whether your company adds other external devices to the computer or has done other uses
5. Check whether the communication line is disconnected or the insulation layer stripped off 4. The vibration test table may be lost or forgot the password caused by the system file or self -modification. And return the computer host to redesign it.