What should I do if the salt fog test box cannot be heated for customers? Make the following solutions for customers:
1. There is no output in the instrument, replacement, replacement, replacementmeter.
2. The heating solid relay is not output, and the solid -state relay is replaced.
3. Water deficiency causes heating shielding, check whether the water level is normal, and whether there is an errors in floating balls.
4. The heating pipe is damaged, check the heating pipe resistance, please replace it if it is abnormal.
5. Super temperature protection misunderstanding, adjust the protection temperature value, if you still cannot solve it, please replace the ultra -temperature protection.