We Have More Than 10 Years of Experience.
   
   
top-ban
  1. HOME > News

Installation conditions sand dust test chamber

2022-12-12

Sand dust test chamber installation conditions
Sand dust test chamber installation site
The ground is flat and well ventilated

There are no strong vibrations around the equipment

There is no strong electromagnetic field around the equipment

No flammable, explosive, corrosive substances and dust around the equipment

Sand dust test chamber Installation requirements

Leave sufficient operating and maintenance space around the equipment as shown in the picture:

A: Not less than 60cm

W: not less than 60cm

C: not less than 120cm

(Note: the inclination angle of the equipment should not exceed 15°C)
Sand dust test chamberInstallation ambient conditions


Temperathours: 5°C~+28°C(average temperature within 24 hours≤ 28°C)

Relative humidity: ≤85%RH

Air pressure: 86kPa~106kPa
Power supply conditions Three-phase four-wire + protective earth wire, Voltage range: AC (380±38) V

Permissible frequency fluctuation range: (50±0.5)Hz

The earth resistance of the protective earth wire is less than 4Ω

span>
Sand dust test chamberrequires the user to use an air or power switch with corresponding capacity for the equipment at the installation site, and this switch must be used independently for this equipment