Thermal shock box with temperature cycle Satisfies the test methods are: SJ/T10187-91Y73 series temperature change test chamber - one-box type; SJ/T10186-91Y73 Series Temperature Change Test Box - Two Box Type; GB/T2423.22-1989 temperature change test; GB/T2423.1- 1989 Low Temperature Test Method; GB/T2423.2-1989 Hochtemperature test method; GJB360.7-87 Temperature Shock Test; GJB150.5-86 Temperature Shock Test; IEC68-2-14_Test method N_Temperature change; GJB367.2-87 405 Temperature Shock Test; GB/T 2423.22-2002 temperature change; GB/T 2424.13-2002 Test Method Temperature Change Test Guideline; EIA 364-32 thermal shock (temperature cycle) test method for electrical connectors and environmental assessment of socket outlets etc.
The thermal cycle performance parameters Shockbox are as follows:
1. Low temperature impact temperature range: A-20 ℃; B-40 ℃; C: 55 ℃; D: -65 ℃ (manufacturer also provides free high temperature up to 150 degrees).
two , temperature deviation: ±2℃.
3. Temperature variation: ±0.5℃.
4 Temperature recovery time: ≤5min.
5. Temperature recovery conditions: high temperature 150℃ exposure for 30 minutes, low temperature -20℃ for 30 minutes, high temperature at 150℃ for 30 minutes, low temperature at -40℃ for 30 minutes, high temperature at 150°C for 30 minutes, low temperature at -55°C for 30 minutes, high temperature at 150°C for 30 minutes, low temperature at -65°C for 30 minutes.
6. Thermal shock transmission method: pneumatic drive.
7. The heating time of the high-temperature storage room: 30min (+25℃~+200℃).
8. The cooling time of the low-temperature storage room: 65 minutes (+25℃~-75℃).
You can Send Message or email info@qinsun-lab.com to us, we will reply tu you within 24 hours.Now tell us your need,there will be more favorable prices!
Home |
Product |
About |
Contact
Email: info@qinsun-lab.com
No. 2578 Gu Dai Road, Shanghai
Copyright © 2021.Qinsun Instruments Co., Ltd. All rights reserved.Sitemap