Atlas SEPAP MHE Environmental Test Chamber

Atlas SEPAP MHE Environmental Test Chamber

The Atlas SEPAP MHE environmental test chamber is used to perform accelerated light aging of polymers in order to understand the aging mechanism of polymers exposed to the outdoors and to check the chemical evolution at the molecular level....

info@qinsun-lab.com
The Atlas SEPAP MHE environmental test chamber is used to perform accelerated light aging of polymers in order to understand the aging mechanism of polymers exposed to the outdoors and to check the chemical evolution at the molecular level. SEPAP MHE uses mercury vapor lamps, which can replicate and accelerate the chemical mechanisms that lead to the gradual loss of polymer performance throughout its service life. SEPAP MHE can operate under irradiation levels of 90W/m2 to 300W/m2 (290 to 420nm).

Features:

Mercury vapor lamp (1000W)
1040cm2 sample rack capacity
Irradiation range 90 to 300W/m2 (290nm to 420nm)
Precise control of irradiance, black mark temperature (BST) and indoor temperature
Radio controlled sample holder lighting and BST monitor XENOSENSIV
Indoor temperature up to 70°C
Atlas
BST up to 100 °C
Full-color multilingual user interface
Relative humidity control using ultrasonic humidifier (SEPAP MHE+)
Relative humidity range 10-95% (SEPAP MHE+)
60L integrated reservoir for sample spraying (SEPAP MHE+)

Applicable standards:

D27 3064
D27 5435
DIN EN 13206
DIN EN 16472
EN 50289-4-17
ISO 10640
NF C 32062-2
NF EN 13207
 

Precautions for xenon lamp aging test machine:

Please keep in mind the precautions of the xenon lamp aging test box, which can avoid the machine malfunction due to improper operation by people, and the correct maintenance method can prolong the life of the machine.
1. Installation location of xenon lamp aging test machine
①. The installation location should consider the heat dissipation and the location where it is easy to check and repair.
② The distance between the wall and any other machines should be more than 600mm.
③. Set on a flat and vibration-free ground.
④. Choose well-ventilated and avoid direct sunlight.
⑤ This machine should be kept away from heat sources and flammable and explosive materials.
⑥ The power supply lines and water supply and drainage pipes should be as short as possible.
⑦. Please choose a place with less dust and moisture and good ventilation for installation
2. Precautions for the use of xenon lamp aging test machine:
① During the operation, please do not open the door of the box unless it is necessary, otherwise it may lead to undesirable consequences.
②. In order to provide the test product of this machine and the safety protection of the operator, please check the water pressure controller and the water level of the circulating water tank regularly.
③. Correctly install the wet bulb test cloth to ensure the correct relative humidity is measured.
④. Please read this manual carefully before operating this machine.
What Can I Do For You?

You can also send a message to us by this email info@qinsun-lab.com, we will reply tu you within 24 hours.Now tell us your need,there will be more favorable prices!

toTop